ao <br />N <br />24'-1 1 /2" T-8" <br />PROPOSED OMEGA EQUIPMENT LAYOUT - AA <br />(ROOM DIMENSIONS SWk,-L BE FIELD VERIFIED) <br />(CONTRACTOR SHALL VERIFY LOCATION OF SUPPORT BEAMS AND JOISTS) <br />(RECOMMENDED FINISHED CEILING HEIGHT: 9'-091 <br />) <br />(DASHED LINES DEPICT EQUIPMENT TRAVEL / SWING; <br />SERVICE CLEARANCES; DOOR SWINGS) <br />EQUIPMENT CLOSET NOTES: <br />EACH EQUIPMENT CABINET HAS A MINIMUM AMOUNT OF SPACE AROUND IT THAT IS REQUIRED TO <br />BE KEPT CLEAR SO THAT SERVICE MIGHT BE PERFORMED AND CABLES / METAL FLEX CONDUIT <br />MAY BE CONNECTED AND DRAPED. EXACT CLEARANCE DIMENSIONS ARE DICTATED BY THE <br />GREATER OF PREVAILING CODE OR THE MANUFACTURER'S SPECIFICATION. REFER TO NEC 110.26, <br />EQUIPMENT DETAILS AND CONSTRUCTION NOTE #6 ON SHEET A-4. <br />THESE DRAWINGS ARE COPYRIGHTED. THEY ARE RESTRICTED TO THE SOLE USE OF <br />GAMMA TECH, INC. <br />THESE DRAWINGS ARE FOR CONCEPTUAL INTERPRETATION ONLY AND ARE NOT <br />INTENDED FOR CONSTRUCTION. <br />THESE DRAWINGS MAY NOT BE COPIED OR DISTRIBUTED TO ANY OTHER INDIVIDUAL(S) <br />OR ORGANIZATION(S) WITHOUT A WRITTEN CONSENT BY GAMMA TECH, INC. <br />OMEGA MEDICAL IMAGING, INC. <br />HEATING, VENTILATING AND AIR CONDITIONING REQUIREMENTS <br />OPERATING TEMPERATURE RANGE: 68' F (20' C) - 75' F (24' C). <br />INTAKE AIR TO ALL EQUIPMENT SHALL BE MAINTAINED WITHIN THIS RANGE. <br />RELATIVE HUMIDITY RANGE: 30% - 70% NON CONDENSING. <br />REFER TO EQUIPMENT SCHEDULE FOR INDIVIDUAL EQUIPMENT BTU/HR RATING. <br />FACILITY: <br />Florida Hospital <br />Zephyrhills, FL <br />DRAWN BY: <br />Dave Moser <br />DATE: <br />01.30.2012 <br />FILE NAME: <br />FHZ122011 <br />SCALE: <br />1 /4" = 1 ' <br />SHEET: <br />B-1 <br />