CT LAB GENERAL NOTES <br />1. REFER TO GE HEALTHCARE DRAWINGS NUMBER/ REFERENCE # 2384867 <br />DATED 04/14/2016 OR LATER FOR CT LAB. CONTRACTOR SHALL REVIEW <br />PHILIPS RADIOLOGY DRAWINGS AS A PART OF THESE CONSTRUCTION <br />DOCUMENTS. DEVICES SHOWN ON GE HEALTHCARE DRAWINGS SHALL <br />TAKE PRECEDENCE IN THE EVENT OF ANY INCONSISTENCIES IN DESIGN <br />DOCUMENTS. FURNISH AND INSTALL WIRING, CONDUIT RACEWAYS, <br />JUNCTION BOXES, CIRCUIT BREAKERS, RELAYS SHUNT TRIP PUSH <br />BUTTONS AND ALL OTHER ELECTRICAL ITEMS NOT PROVIDED BY PHILIPS. <br />PULL ALL CABLES FURNISHED BY PHILIPS AND MAKE ALL CONNECTIONS <br />AS DIRECTED BY GE HEALTHCARE. <br />2. WIRES SHALL BE AWG STRANDED, FLEXIBLE, THERMO-PLASTIC <br />COVERED, COLOR CODED COPPER ONLY. TYPE THW 600V WITH <br />MULTIPLE STRANDS. WIRES SHALL BE CUT TEN FOOT LONG AT <br />OUTLET BOXES, DUCT TERMINATION POINTS OR STUBBED CONDUIT <br />ENDS, UNLESS OTHERWISE SPECIFIED. ALL CONDUCTORS, POWER, <br />SIGNAL, AND GROUND MUST BE RUN IN CONDUIT OR DUCT SYSTEM <br />(SEPARATED BY BARRIERS). CONTRACTOR SHALL RING OUT AND <br />TAG ALL WIRES AT BOTH ENDS. WIRE RUNS MUST BE CONTINUOUS <br />COPPER AND FREE FROM SPLICES. <br />3. COORDINATE WITH RADIOLOGY VENDOR DRAWINGS AND PROVIDE <br />NON-FERROUS ELECTRICAL DUCT WITH NON-FERROUS DIVIDERS OR <br />INDIVIDUAL CONDUIT TO PROVIDE SEPARATION FOR THE FOLLOWING <br />CABLE TYPES: <br />(1) 24VOLT <br />(2) 300 VOLT AND LESS <br />(3) 301-600 VOLTS <br />(4) HIGH TENSION X-RAY CABLES <br />(5) VIDEO CABLES <br />ALL CABLES SHALL BE SEPARATED FOR THEIR ENTIRE RUN. <br />4. EXACT LOCATIONS OF ALL WIREWAYS, CONDUIT RUNS AND JUNCTION <br />BOXES SHALL BE IN ACCORDANCE AND COORDINATED WITH THE <br />PHILIPS VENDOR DRAWINGS. <br />5. FOR CONDUIT REQUIRED BETWEEN EQUIPMENT AND SIZES SEE <br />PHILIPS DRAWINGS. <br />6. SEE PHILIPS DRAWINGS FOR THE EXACT LOCATION AND MOUNTING <br />HEIGHTS OF ALL NEW ELECTRICAL DEVICES TO BE INSTALLED. <br />7. EACH CONDUIT SHALL TERMINATE IN PLASTIC BUSHING. LEAVE 10' <br />WIRE PIGTAIL AT ALL TERMINATION POINTS. <br />8. GE HEALTHCARE SYSTEMS REPRESENTATIVE SHALL PERFORM A <br />LEAKAGE CURRENT TEST PER NFPA 99 CHAPTER 10-3.3. ON ALL <br />PHILIPS EQUIPMENT INSTALLED UNDER THE PROJECT SCOPE. UPON <br />COMPLETION OF THE LEAKAGE TEST THE MANUFACTURER'S <br />REPRESENTATIVE SHALL SUBMITT THE RESULTS TO THE ELECTRICAL <br />ENGINEER OF RECORD FOR REVIEW AND OR APPROVAL. THE FINAL <br />OFFICIAL RESULTS OF THE TEST SHALL BE PROVIDED TO THE <br />OWNER'S REPRESENTATIVE, GENERAL CONTRACTOR, AND ELECTRICAL <br />ENGINEER OF RECORD PRIOR TO THE FINAL INSPECTION. <br />9. ELECTRICAL CONTRACTOR SHALL PROVIDE CODE COMPLIANT <br />PENETRATIONS FOR ALL NEW CONDUITS ENTERING/EXITING <br />FIRE RATED WALLS. ALL EXISTING NON -COMPLIANT ELECTRICAL <br />PENETRATIONS IN THE PROJECT SCOPE SHALL_ BE THE <br />RESPONSIBILITY OF THE CONTRACTOR TO CORRECT. <br />NOTE: ONLY ELECTRICAL WIRING SHALL BE IN ELECTRICAL DUCT. <br />NO COOLING OR OTHER NON -ELECTRICAL CONNECTIONS SHALL <br />BE ROUTED IN SAME DUCT. <br />GROUNDING NOTES: <br />1. GROUNDING SHALL COMPLY WITH NEC ARTICLE 517 AND NFPA <br />99. ALL EQUIPMENT SHALL HAVE REDUNDANT GROUND. <br />2. ALL NIPPLES REQUIRE GROUND STUDS. <br />3. ALL DUCT WORK REQUIRES GROUND JUMPERS AT ALL JOINTS. <br />4. ALL GE CABINETS MUST BE CONNECTED TO THE DUCT OR JUNCTION BOXES <br />WITH DIRECT BONDING OR JUMPER WIRES. <br />5. OVERHEAD TUBE OR TV MONITORS MUST HAVE A GROUND WIRE <br />FROM THE DUCT OR JUNCTION BOX TO THE TV MONITOR OR <br />X-RAY TUBE UNIT. <br />RL <br />l <br />C <br />LI <br />3 <br />DS <br />TO CT EQUIP <br />ELECTRICAL PARTIAL <br />LIGHTING FLOOR PLAN - NEW CONSTRUCTION <br />SCALE: 1/4" = 1'- 0" <br />LA-22,24,26 <br />EXISTING EQUIP <br />CONDUITS ROUTE UPS <br />6 <br />LB-31 <br />I <br />277/480V <br />SERVICE <br />CU-1 <br />of <br />EXISTING <br />5 `ITON CU <br />480/277V <br />WP <br />8 <br />LB-31 LASER L27 <br />LB-27 <br />LB-33 =0 LB-29 <br />IQ <br />"® LB-29 ; <br />LB-29 <br />B-13 <br />O ❑� <br />6 <br />5 <br />j 'EXISTING <br />W j MONITOR <br />LB-31(900 <br />LA-21,23,25 <br />82S <br />LA-41 <br />EXISTING <br />DIETER <br />ELECTRICAL PARTIAL <br />POWER & SYSTEMS FLOOR PLAN - NEW CONSTRUCTION <br />SCALE: 1/4" = 1'- 0" <br />1) ELECTRICAL CONTRACTOR SHALL PROVIDE DEDICATED NEUTRALS FOR EACH NEW CIRCUIT AND <br />MODIFIED EXISTING CIRCUIT WITHIN THE PROJECT SCOPE. MULTI -CIRCUIT APPLICATION OF <br />SHARED NEUTRALS SHALL NOT BE ALLOWED. <br />2) CONTRACTOR SHALL CLEAN, REPLACE ALL INOPERABLE BALLAST AND RELAMP ALL EXISTING <br />LIGHTING FIXTURES BEING REUSED IN RENOVATION. REPLACE ALL LAMPS IN EXISTING LIGHTING <br />FIXTURES WITH SAME COLOR TEMPLATE AND COLOR RENDERING INDEX TO ESTABLISH UNIFORM <br />DISTRIBUTION AND CONSISTENT VISIBLE APPEARANCE. RECONNECT ALL EXISTING LIGHTING <br />FIXTURES AS SHOWN. <br />3) RL DENOTES RELOCATED DEVICE. <br />KEY NOTES <br />O1 PROVIDE NEW CEILING MOUNTED LIGHTING FIXTURES AND RECONNECT TO EXISTING CIRCUIT <br />MADE AVAILABLE BY DEMOLITION. SEE LIGHTING FIXTURE SCHEDULE THIS SHEET FOR <br />ADDITIONAL INFORMATION. <br />0 EXISTING LIGHTING SWITCH SHALL BE REWORK TO ACCOMMODATE NEW LOCATION. CONNECT <br />TO EXISTING CIRCUIT MADE AVAILABLE BY DEMOLITION AND PROVIDE DEDICATED NEUTRAL FOR <br />EACH CIRCUIT REUSED. <br />OPROVIDE NEW'XRAY IN USE' LIGHT AND DOOR CONTROL DEVICE. CIRCUIT TO BE EXTENDED TO NEW <br />VENDOR EQUIPMENT. COORDINATE WITH VENDOR'S DRAWINGS FOR INSTALLATION DETAILS, <br />QUANTITIES, AND LOCATIONS WITHIN DOOR JAMB FRAME. <br />OCONTRACTOR SHALL PROVIDE NEW 48OV, 3PH, 4W, CIRCUIT BREAKER LABELED "ECB-S". PROVIDE <br />SHUNT TRIP ACCESSORY AND TIE TO ENCLOSED CIRCUIT BREAKER. REFER TO RISER DIAGRAM & <br />VENDOR DRAWINGS FOR EXACT CIRCUIT BREAKER, SHUNT -TRIP, ETC REQUIREMENTS AND FOR <br />ALL CONNECTIONS TO THE EQUIPMENT. <br />OEMERGENCY POWER OFF PUSH BUTTON SHALL BE CONNECTED TO ENCLOSED SHUNT TRIP <br />CIRCUIT BREAKER. COORDINATE WITH VENDOR'S DRAWINGS FOR INSTALLATION DETAILS, <br />PUSHBUTTON TYPE, AND FINAL LOCATIONS. <br />ONEW FIRE ALARM DEVICES TO BE CONNECTED TO EXISTING FIRE ALARM SYSTEM SERVING AREA. <br />CONTRACTOR SHALL VERIFY NEW DEVICE IS FULLY FUNCTIONAL. <br />O7 PROVIDE POWER SUPPLY BOX FOR VENDOR EQUIPMENT. REFER TO VENDOR DRAWINGS FOR <br />ADDITIONAL INFORMATION. <br />�8 NEW 480/277V, NEMA 3R SERVICE ENTRANCE DISCONNECT SAFETY SWITCH FOR NEW CT <br />EQUIPMENT. SEE RISER DIAGRAM FOR ADDITIONAL INFORMATION. <br />O9 PROVIDE 4 #8, 1 #10G IN 1 INCH CONDUIT FOR NEW 120/208V, 3P, 60AMP DISCONNECT <br />SERVING NEW MECHANICAL EQUIPMENT FROM EXISTING PANEL LA. REFER TO MECHANICAL <br />PLANS AND PANEL SCHEDULES FOR ADDITIONAL INFORMATION. <br />10 PROVIDE ]-BOXES PER VENDOR EQUIPMENT. REFER TO VENDOR DRAWINGS FOR ADDITIONAL <br />INFORMATION. <br />IF-- LIGHTING FIXTURE SCHEDULE 11 <br />� ITYPE MANUFACTURER CATALOG NO, LAMP/LUMENS VOLT I MOUNTING REMARKS I <br />LITHONIA <br />2X2 COMPACT FLUORESCENT <br />F1 <br />LIGHTING <br />2SPG217A12125MVOLTI/3 FLUORESCENT <br />WXIABWW-OF-X1U <br />UNV <br />GRID STRAIGHT AND U LAMPS <br />X-RAY ON WARNING LIGHT/ <br />XA <br />GE <br />WARNING CONTROLLER <br />E4502RL OR EQUAL <br />GE-800-279-7925 <br />RENOVATION LEGEND <br />SYMBOL <br />DESCRIPTION <br />NEW CONSTRUCTION <br />- - <br />EXISTING TO REMAIN <br />CONNECT TO EXISTING AT THIS POINT <br />S�itcc:), ILILC <br />AA 26001479 <br />Sheet Notes: <br />Smith <br />Seckman. <br />Reid, Inc. <br />301 North Cattlemen Road, Suite 300 <br />Sarasota, FL 34232 <br />(941) 907-7750 <br />FAX: (941) 907-7751 <br />www.ssr-iiic.com <br />Firm Certification - CA-0000857 <br />SSR# 1636044.0 <br />Erick Phelps - PE #64598 <br />Christopher E. Durham - PE 976655 <br />No. 76655 <br />_ -a <br />C1 °°° STATE OF <br />NAU <br />Christopher Edward Durham <br />PE license No. 76655 <br />Key Plan <br />MARK ISSUE DATE DESCRIPTION <br />Issue Log <br />CONSTRUCTION <br />DOCUMENTS <br />FLORIDA CANCER SPECIALISTS <br />EQUIPMENT REPLACEMENT <br />38010 MEDICAL CENTER AVE. <br />ZEPHYRHILLS, FL 33540 <br />Drawing Title <br />ELECTRICAL <br />PARTIAL FLOOR PLAN <br />- NEW CONSTRUCTION <br />71604 <br />06/29/2016 <br />Scale 1/4" = 1' - 0" <br />Sheet Number <br />E301 <br />Drawn By KDS <br />Approved By MLG <br />Filename <br />SaltCo, LLC 2007 <br />