Laserfiche WebLink
<br />SIGMA 34 & 40 INSTALLATION, TEST AND <br />ACCEPTANCEPROCEDVRE <br />ITA-34-902130 Rev J <br /> <br />Perry Baromedical Corporation <br /> <br />1.0 INTRODUCTION <br /> <br />1.1 PURPOSE <br />The purpose of this procedure is to establish the requirements for SIGMA 34 and/or <br />SIGMA 40 installation, final checkout of all systems, and criteria for customer acceptance. <br /> <br />1.2 SCOPE <br />The scope of this specification covers the following areas: <br /> <br />a. SIGMA 34 and/or SIGMA 40 on-site installation and interface with customer's <br />facility <br /> <br />b. Functional checkout of chamber and all associated systems. <br /> <br />c. Final acceptance by customer. <br /> <br />1.3 PREREQUISITES <br />The following items shall have been completed before shipment to customer's location. <br /> <br />a. Factory acceptance testing to Perry Baromedical Corporation Specifications. <br /> <br />b. Customer's representative shall be notified at least one (1) day in advance of the <br />scheduled performance of any or all tests required by this procedure. <br /> <br />c. Building service requirements (oxygen. air, exhaust, and electrical grounding) <br />installed by customer. <br /> <br />2.0 INSTALLATION <br /> <br />2.1 Assemble the chamber on the location as directed / given by the customer. <br />2.2 Lock wheel brakes to prevent chamber movement. <br />2.3 With the chamber assembly located and secured in position, install supply, exhaust, <br />electrical power and ground interconnects between the chamber and facility connection. <br />2.4 Check following torque values: Tie rod nuts- 90 ft. lbs. for SIGMA 34 <br />Tie rod nuts- 110ft. Ibs. for SIGMA 40 <br />Pivoting door bolts- 175 ft. Ibs. for both SIGMA models. <br /> <br />3.0 INTERCONNECT PRESSURE CHECK <br /> <br />NQIE: <br /> <br />A customer representative along with an individual from Perry Baromedical <br />Corporation must be present to witness the tests specified in Sections 3.0 and 4.0. <br />The completion of these tests and the results accepted by the customer signifies <br />acceptance and final turnover of the system to the customer. <br /> <br />Notice: The only controlled copies of this dOCUMent as-. the electronic read-only verslema located on np' drive, <br />and the .igned master copy maintained by the Quality deparbnenl The document us.r Is ....ponslb.. for verifying that the latest <br />v....ion Is availabl. for use, by cMc:klng the electronic dlractoty to detennin. the cumtnt ntvlaion level before using Level 3 or <br />, -..-. · ..--..-.... Page 3 of 9 <br />